Bacteria and microorganism rapid detector
- TIANWEI
- CHINA
- 7-14 DAYS AFTER PO
Practicability —— The upper and lower limits can be set according to the requirements of environmental testing, so that data can be quickly evaluated and early warning, and the surface cleanliness can be quickly screened.
High sensitivity —— 10-15~10-18 mol
Fast speed-the conventional culture method is more than 18-24h, while ATP only takes more than ten seconds.
Product introduction:
The device is a brand new upgraded product with a large touch screen instead of traditional buttons. The operation uses a biochemical reaction method to detect ATP content. The ATP fluorescence detector is based on the principle of firefly luminescence and uses the "luciferase-luciferin system" to quickly detect adenosine triphosphate (ATP). The ATP swab contains a reagent that can lyse the cell membrane, which can release ATP from the cell, react with the specific enzyme contained in the reagent to produce light, and then use a fluorometer to detect the luminescence value. The number of microorganisms is proportional to the luminescence value. Since all living organisms contain a constant amount of ATP, the ATP content can clearly indicate the residual amount of microorganisms and other organisms in the sample, which is used to judge the sanitary condition.
Instrument characteristics:
Practicability —— The upper and lower limits can be set according to the requirements of environmental testing, so that data can be quickly evaluated and early warning, and the surface cleanliness can be quickly screened.
High sensitivity —— 10-15~10-18 mol
Fast speed-the conventional culture method is more than 18-24h, while ATP only takes more than ten seconds.
Feasibility-There is a clear correlation between the number of microorganisms and the ATP contained in the microorganisms. By detecting the ATP content, the number of microorganisms in the reaction can be obtained indirectly
Operability —— The traditional culture method needs to be operated by trained technicians in the laboratory; and the ATP rapid cleanliness detection operation is very simple, and only simple training can be performed by general staff on-site.
Better experience-The test tube adopts a pluggable and flexible design, which can be cleaned regularly and used for a long time to extend the life of the instrument.
The main parameters:
1. Display: 3.5-inch high-precision graphic touch screen
2. Processor: 32-bit high-speed data processing chip
3. Detection accuracy: 1×10-18mol
4. Coliform group: 1-106cfu
5. Detection range: 0 to 9999 RLUs
6. Detection time: 15 seconds
7. Detection interference: ±5﹪ or ±5 RLUs
8. Operating temperature range: 5℃ to 40℃
9. Operating humidity range: 20—85﹪
10. ATP recovery rate: 90-110%
11. Detection mode: RLU, coliform screening
12. 50 user ID settings
13. The number of result limits that can be set: 251
14. Automatically judge qualified and unqualified
15. Automatic statistics of pass rate
16. Built-in self-calibration light source
17. Self-check after booting in 30 seconds
18. Equipped with a miniUSB interface, the results can be uploaded to the PC
19. Equipped with dedicated software driver U disk instead of traditional CD
20. Instrument size (W×H×D): 188 mm×77mm×37mm
21. Use rechargeable lithium battery without battery replacement
22. Standby state (20℃): 6 months
23. Chinese operation manual
24. Stable liquid luciferase
25. Wetted integrated collection swab
Application:
The ATP fluorescence detector has a wide range of uses and can be used in various industries such as food, medicine and health, medicine, daily chemical, papermaking, industrial water treatment, national defense and environmental protection, water administration, customs entry and exit quarantine and other law enforcement agencies.